301-1 Tap Density Tester Calibration
TAP DENSITY TESTER CALIBRATION
Speaking of tap density tester calibration, HMKTest usually take following calibration procedures for LABULK 0335 Intelligent Tap Density Tester:
1. Count and check rpm i.e. revolutions per minute as per international standards such as ASTM B527, ISO 3953 LABULK 0335 Tap Density Tester meets all the standards about tap density measurement. rpm requirement is mandatory in pharma testing but not for metallic powders, for which rpm is seldom regulated.
2. Measure and check the drop height according to international standards, LABULK 0335 Tap Density Tester meets all international standards.
3. Other procedures specified by relevant international standards. Such as in pharmaceutical test, weigh the tapping assembly including the cylinder and cylinder holder and check if weight is in the right range.